REV. A 1 OF 4
Replacing the 2424 Nebulizer
The pneumatically controlled nebulizer converts solvent flow to an aerosol. There are two
nebulizer configurations:
• High-flow for solvent flows of 0.3 to 3.0 mL/min
• Low-flow for solvent flows of 0.005 to 0.5 mL/min
When to Replace
• If the nebulizer is damaged
• If your application requires a higher or lower flow rate
Parts Required
• High flow nebulizer, # 205000458
• Low flow nebulizer, # 205000459
Tools Required
• 5/16-inch open-end wrench
Preparation
1. Before performing this procedure, stop solvent flow to the system, then turn off gas
flow.
2. Always wear powder-free gloves or finger cots when handling the nebulizer to
prevent contact with residual solvents.
Replacing the 2424 Nebulizer
REV. A 2 OF 4
Removing the Old Nebulizer
1. Turn the detector off, then disconnect the power cord.
CAUTION: IF YOU USED THE NEBULIZER HEATER, WAIT 30 MINUTES FOR THE NEBULIZER
TO COOL BEFORE TOUCHING IT.
2. Remove the left front cover.
3. If a column or second detector is connected to the system, disconnect the solvent
inlet tubing from the front of the nebulizer as follows:
a. Use the 5/16-inch wrench to loosen the compression screw that holds the inlet
tubing in place.
Figure 1 - Removing the Compression Screw
b. Remove the solvent inlet tubing from nebulizer.
4. Push in the quick-disconnect tubing fitting on the right-hand side of the nebulizer, and
pull out the gas inlet tubing.
Figure 2 - Removing the Gas Inlet Tubing
Replacing the 2424 Nebulizer
REV. A
您可能关注的文档
- 3355 TQS 超高效液相色谱串联质谱 2014-5-9.docx
- NuGenesis LMS安装和配置指南.pdf
- QDa 检测器概述和维护指南.pdf
- 沃特世仪器说明书沃特世仪器说明书masslynx安装指南.doc
- 沃特世仪器说明书QDa QC应用文集.pdf
- 沃特世仪器说明书Replace 2424 Lamp.pdf
- 沃特世仪器说明书Rebuild 2998 flow cell.pdf
- 沃特世仪器说明书Plunger Seal Wash.pdf
- 沃特世仪器说明书Performance Maintenance.pdf
- I沃特世仪器说明书ntro.pdf
- GB/T 29324-2024架空导线用碳纤维增强复合材料芯.pdf
- 《GB/T 29324-2024架空导线用碳纤维增强复合材料芯》.pdf
- GB/T 43905.1-2024焊接及相关工艺中烟尘和气体取样的实验室方法 第1部分:电弧焊中烟尘排放速率的测定和分析用烟尘的收集.pdf
- 《GB/T 43905.1-2024焊接及相关工艺中烟尘和气体取样的实验室方法 第1部分:电弧焊中烟尘排放速率的测定和分析用烟尘的收集》.pdf
- 中国国家标准 GB/T 43905.1-2024焊接及相关工艺中烟尘和气体取样的实验室方法 第1部分:电弧焊中烟尘排放速率的测定和分析用烟尘的收集.pdf
- 中国国家标准 GB/T 18910.21-2024液晶显示器件 第2-1部分:无源矩阵单色液晶显示模块 空白详细规范.pdf
- GB/T 18910.21-2024液晶显示器件 第2-1部分:无源矩阵单色液晶显示模块 空白详细规范.pdf
- 《GB/T 18910.21-2024液晶显示器件 第2-1部分:无源矩阵单色液晶显示模块 空白详细规范》.pdf
- GB/T 43860.1220-2024触摸和交互显示 第12-20部分:触摸显示测试方法 多点触摸性能.pdf
- 中国国家标准 GB/T 43860.1220-2024触摸和交互显示 第12-20部分:触摸显示测试方法 多点触摸性能.pdf
文档评论(0)