Waters Alliance HPLC Systems:
Guidelines for Ensuring
Optimum Performance
Performance Maintenance
To ensure uninterrupted operation of your 2695 or 2795 Separations Module, we
recommend that you or your Waters field service engineer complete performance
maintenance (PM) once a year. Scheduled, pre-failure PM is a cost-effective way to
ensure accurate, reproducible results and the highest level of reliability.
All the normal-wear and maintenance parts that are necessary to perform annual
scheduled maintenance are included in the PM Kit (see table).
PM Kit Contents
Description Part Number
2690/2695 PM Kit WAT270944
2790/2795 PM Kit 201000107
Performance maintenance consists of the replacement of the following parts:
• Seal-wash seals
• Plunger and plunger seals
• Inlet check valves and cartridges
• Inline filter
• Injection needle and seals
• Syringe
NOTE: If you use the system continuously (for example, during nights and
weekends), or if you use aggressive solvents such as buffers, you may need to
perform PM more frequently.